TB 1-4920-443-35
(l) With the tachometer tester, set RPM selector to
the remaining speeds listed below. At
each
tachometer tester speed, verify that the TI calculating counter and TI indicator indicates within the
tolerance limits listed below. Record results.
Table 48-4 TI Calculating Counter vs. TI Indicator
Tachometer
TI Calculating
TI Indicator TI Indicator (%)
AEDATS
Tester Speed
Counter
Digital .3%
Analog .3%
.1 Hz
(%)
.3%
RPM3
40 (28 Hz)
39.9 40.1
60 (42 Hz)
59.9 60.1
80 (56 Hz)
79.9 80.1
100 (70 Hz)
99.9 100.1
NOTE
For AEDATS II Channel names, refer to AEDATS (H345-1) Technical Manual, Appendix C and
for AEDATS IV Channel names, refer to AEDATS (H355-7) Technical Manual. Chapter 6.
AEDATS Alignment RPM3 - See Calibration of Automatic Data Acquisition H345-1
(AEDATS II), or H355-7 (AEDATS IV) for FEDS Alignment requirements.
(m) Set the tachometer tester to
10%, the motor direction switch to
STOP, and the power switch to
OFF.
49. T53/T55/T63/T64 Gas Generator (N1) Speed Indicator and T64 Gas Generator (Ng) Speed Indicator
a. Performance Check
(1) Connect the tachometer generator test set to
pins A
and B of
engine harness at
Tachometer (N1).
(2) Set the TI calculating counter controls (Channel 1) as follows:
Table 49-1 TI Calculating Counter Controls
CH1 DEC
5
CH1 C
14286
N
CH1 10 Period
0
CH1 Auto
depressed
CH1 x10
released
channel 1
depressed
SAMPLE RATE (FAST)
1
TEST
released
(3) Repeat steps h through m in paragraph 50
(4) With the tachometer tester, set RPM selector to
the remaining speeds listed below. At
each
tachometer tester speed, verify that the TI calculating counter and TI indicator indicates within the
tolerance limits listed below. Record results.
Table 49-2 TI Calculating Counter vs. TI Indicator
Tachometer
TI Calculating
TI Indicator TI Indicator (%)
AEDATS
Tester Speed
Counter
Digital .3%
Analog
.1 % NG
(%)
.3%
.3%
40 (28 Hz)
60 (42 Hz)
80 (56 Hz)
100 (70 Hz)