TM 5-6675-323-14
Section III OPERATOR MAINTENANCE
LUBRICATION INSTRUCTIONS.
5-8.
NOTE
These lubrication instructions are mandatory.
5-8.1 Lubricate Microscan. Clean and lubricate bearing surfaces monthly with
b e a r i n g c l e a n e r ( I t e m 5 , A p p e n d i x E) and general purpose oil, (Item 14, Appendix E)
as follows:
Death or serious injury may occur from electrical shock unless power
cord is unplugged before servicing.
U n p l u g power cord.
a.
b. Wipe all exposed surfaces to remove dirt from microscan.
L o o s e n X- and Y-axis thumbscrews.
c.
d.
Clean bearing surfaces with bearing cleaner.
Apply small amount of oil to bearing surfaces.
e.
f. Move carriage in X- and Y-directions to distribute oil evenly.
Clean optical mount track with bearing cleaner.
9"
h. Apply small amount of oil to track on optical mount.
Raise and lower optical mount to distribute oil by turning focusing knobs.
i.
Clean spindles with bearing cleaner.
j.
k. Apply small amount of oil to spindles.
Rotate
spindles
to
distribute
oil.
1.
Wipe all surfaces clean of excess oil.
m.
Reclean all surfaces that contact film.
n.
5-23