TB 1-4920-443-35
(13) MT13:
0-50 PSIG (T53/T55/T63/T64)
Using the pressure calibrator, apply the following pressures to MT13 and record the results.
Table 43-13 MT13 Alignment
Target
Calibrator
Vertical Scale
AEDATS
Pressure (psig)
Pressure (psig)
Bottom Row #11
Pdyn_sup ( 1.0 PSI)
Indicator ( 1.0 PSI)
See Note below
VENT
10
20
30
40
50
RCAL=
NOTE
For AEDATS II Channel names, refer to AEDATS (H345-1) Technical Manual, Appendix C and
for AEDATS IV Channel names, refer to AEDATS (H355-7) Technical Manual. Chapter 6.
AEDATS Alignment MT13 - See Calibration of Automatic Data Acquisition H345-1
(AEDATS II), or H355-7 (AEDATS IV) for FEDS Alignment requirements.
a. Performance Check
Ensure that the system is in the T700 mode (Reference FEDS operator's manual).
(1) CDP Transducer - 0-600 In Hg Abs (T700 & T64)
Using the pressure calibrator, apply the following pressures to the CDP transducer located in
Cabinet 11.
Table 44-1 CDP Transducer Alignment
Target Absolute
Calibrator Absolute
Vertical Scale
AEDATS
Pressure (In Hg)
Pressure (PSIA)
Bottom Row #7
CDP 0.3 In Hg
Indicator ( 0.3 In Hg)
See Note below
ARO
150
73.7
300
147
450
221
600
294.7
NOTE
For AEDATS II Channel names, refer to AEDATS (H345-1) Technical Manual, Appendix C and
for AEDATS IV Channel names, refer to AEDATS (H355-7) Technical Manual. Chapter 6.
AEDATS Alignment CDP - See Calibration of Automatic Data Acquisition H345-1
(AEDATS II), or H355-7 (AEDATS IV) for FEDS Alignment requirements.
a. Performance Check
Ensure that the FEDS is in the T53/T55/T63/T64 mode (Reference FEDS operator's manual).
(1) MT1:
0-100 PSIG (T53/T55 & 714A)
Using the pressure calibrator, apply the following pressures to MT1 and record the results.