TB 1-4920-443-35
(9) MT10:
0-1000 PSIG (T53, T55,& 714A)
Using the pressure calibrator, apply the following pressure to MT10 and record the results.
Table 43-9 MT10 Alignment
Target
Calibrator
Vertical Scale
AEDATS
Pressure (psig)
Pressure (psig)
Top Row #10
( 2.0 PSI)
Indicator ( 2.0 PSI)
See Note below
VENT
200
400
600
800
1000
RCAL=
NOTE
For AEDATS II Channel names, refer to AEDATS (H345-1) Technical Manual, Appendix C and
for AEDATS IV Channel names, refer to AEDATS (H355-7) Technical Manual. Chapter 6.
AEDATS Alignment MT10 - See Calibration of Automatic Data Acquisition H345-1
(AEDATS II), or H355-7 (AEDATS IV) for FEDS Alignment requirements.
(10)MT10:
0-1500 PSIG (T64)
Using the pressure calibrator, apply the following pressure to MT10 for T64 and record results.
Table 43-10 MT10 - 0-1500 PSIG - Alignment
Target
Calibrator
Vertical Scale
AEDATS
Pressure (psig) Pressure (psig)
Top Row #10
PFD (2.0 1 PSI)
Indicator ( 2.0 PSI) See Note below
VENT
300
600
900
1200
1500
RCAL=
NOTE
For AEDATS II Channel names, refer to AEDATS (H345-1) Technical Manual, Appendix C and
for AEDATS IV Channel names, refer to AEDATS (H355-7) Technical Manual. Chapter 6.
AEDATS Alignment MT10 - See Calibration of Automatic Data Acquisition H345-1
(AEDATS II), or H355-7 (AEDATS IV) for FEDS Alignment requirements.
40