TB 1-4920-443-35
(5) MT7:
0-200 PSIG (T53, T55, 714A, T63)
Using the pressure calibrator, apply the following pressures to MT7 and record the results.
Table 43-5 MT7 Alignment
Target
Calibrator
Vertical Scale
AEDATS
Pressure (psig)
Pressure (psig)
Top Row #6
( 1 PSI)
Indicator ( 2 PSI)
See Note below
VENT
40
80
120
160
200
RCAL=
NOTE
For AEDATS II Channel names, refer to AEDATS (H345-1) Technical Manual, Appendix C and
for AEDATS IV Channel names, refer to AEDATS (H355-7) Technical Manual. Chapter 6.
AEDATS Alignment MT7 - See Calibration of Automatic Data Acquisition H345-1
(AEDATS II), or H355-7 (AEDATS IV) for FEDS Alignment requirements.
(6) MT8:
0-200 PSIG (T53)
Using the pressure calibrator, apply the following pressures to MT8 for T53 and record the results.
Table 43-6 MT8 Alignment
Target
Calibrator
Vertical Scale
AEDATS
Pressure (psig)
Pressure (psig)
Top Row #8
PTbp ( 1 PSI)
Indicator ( 2.0 PSI)
See Note below
VENT
40
80
120
160
200
RCAL=
NOTE
AEDATS channel name is T53 - PTbp for T53 engines only.
AEDATS Alignment MT8 - See Calibration of Automatic Data Acquisition H345-1
(AEDATS II), or H355-7 (AEDATS IV) for FEDS Alignment requirements.