TB 1-4920-443-35
NOTE
For AEDATS II Channel names, refer to AEDATS (H345-1) Technical Manual, Appendix C and
for AEDATS IV Channel names, refer to AEDATS (H355-7) Technical Manual. Chapter 6.
AEDATS Alignment PT102, PT103, PT104 - See Calibration of Automatic Data Acquisition
H345-1 (AEDATS II), or H355-7 (AEDATS IV) for FEDS Alignment requirements.
(b) Locate PT102 on the Bell mouth and Barometer Panel in Cabinet 10. Connect the low pressure
calibrator directly to the transducer. Apply the following pressures and record the results.
Table 45-5 PT102 Alignment
Pressure (In H2O)
Calibrator Pressure
Vertical Scale
0.25 In H2O
Bottom Row #6
AEDATS
Indicator ( 0.25)
See Note below
VENT
10
20
30
40
50
NOTE
For AEDATS II Channel names, refer to AEDATS (H345-1) Technical Manual, Appendix C and
for AEDATS IV Channel names, refer to AEDATS (H355-7) Technical Manual. Chapter 6.
AEDATS Alignment PT102, PT103, PT104 - See Calibration of Automatic Data Acquisition
H345-1 (AEDATS II), or H355-7 (AEDATS IV) for FEDS Alignment requirements.
(c) Locate PT103 on the Bell mouth and Barometer Panel in Cabinet 10. Connect the low pressure
calibrator directly to the transducer. Apply the following pressures and record the results.
Table 45-6 PT103 Alignment
Pressure (In H2O)
Calibrator Pressure
Vertical Scale
PS12 0.25 In H2O
Bottom Row #7
AEDATS
Indicator ( 0.25)
See Note below
VENT
10
20
30
40
50
NOTE
For AEDATS II Channel names, refer to AEDATS (H345-1) Technical Manual, Appendix C and
for AEDATS IV Channel names, refer to AEDATS (H355-7) Technical Manual. Chapter 6.
AEDATS Alignment PT102, PT103, PT104 - See Calibration of Automatic Data Acquisition
H345-1 (AEDATS II), or H355-7 (AEDATS IV) for FEDS Alignment requirements.
(d) Locate PT101 on the Bell mouth and Barometer Panel in Cabinet 10. Connect the low pressure
calibrator directly to the transducer. Apply the following pressures and record the results.
45