TB 1-4920-443-35
29. Type-J Thermocouple Alignment - T53/T55/T63/T64 Trailer
a. Performance Check
(1) In this section we will align the Type-J thermocouple channels monitoring the T53/T55/T63/T64
engine tests. These channels are listed in the table below.
Table 29-1
Channel Name
Connector
Doric
AEDATS
Temperature
Label
Channel
Channel
Range
* See NOTE below
0-300 oF
Eng oil inlet temp
P801
11
Toil_in
0-300 oF
Eng oil out temp
P802
12
Toil_out
0-120 oF
Bellmouth Temp #1
P803
13
Tt11
0-120 oF
Bellmouth Temp #2
P804
14
Tt12
0-800 oF
15
T31
0-120 oF
Bellmouth Temp #3
P806
16
Twf
0-800 oF
17
T32
0-800 oF
P.T. Bearing Scavenge
P808
18
Tptso
0-800 oF
#2 Bearing Scavenge
P809
19
T2bso
0-300 oF
Dyno-lube Inlet
J20
21
Tdyn tnk
0-300 oF
Dyno-lube FWD
J21
22
Tdyn_FWD
0-300 oF
Dyno-lube AFT
J22
23
Tdyn_AFT
0-300 oF
Dyno FWD/AFT Avg
J23
none
Tdyn_FA
*NOTE: For AEDATS II Channel
names, refer to AEDATS (H345-1) Technical Manual,
Appendix C and for AEDATS IV
Channel names, refer to AEDATS (H355-7) Technical
Manual, Chapter 6.
(2) Access cables 20090768-1 and 20090769-1 on the T53/T55/T63/T64 side of the J-box. These
cables contain the connectors where the thermocouple calibrator will be installed.
(3) As a
preliminary continuity check, install thermocouples at
each connector and verify that the
indicated temperature for the respective channel is approximately the ambient temperature (this test
is optional).
(4) For each temperature channel listed in the following tables, perform steps (1) through (4) below.
(a) Connect the thermocouple (TC) calibrator to the specified connector using J-type TC leads.
(b) Adjust the calibrator controls to obtain each specified temperature setting.
(c) At each temperature set point, record the indicated temperature and verify that the recorded
temperature is within the given tolerance limits.
(d) Calibrate the corresponding AEDATS channel in accordance with AEDATS II (H345-1)
Technical Manual, or AEDATS IV (H355-7) Technical Manual.
NOTE
The Limit Control Panel, Temperature and Speed Limit Control (Steps 35, 36, 37, 39, and
40) may be calibrated in conjunction with the channels marked by "*" in the following
table.