TB 1-4920-443-35
Table 29-2 Continued
Parameter
Connector
DORIC
AEDATS
Setpoint
DORIC
DORIC
AEDATS
AEDATS
Name
Channel
Channel
Temp (F)
Tolerance
Reading
Tolerance
Reading
Limits
Limits (1)
(2)
* Dyno-lube
J20 of dyno
21
Tdyn_tnk
32
30 34
31 to 33
Inlet Temp.
100
98 102
99 to 101
200
198 202
199 to 201
300
298 302
299 to 301
* Dyno-lube
J21 of dyno
22
Tdyn_fwd
32
30 34
31 to 33
Fwd Temp.
100
98 102
99 to 101
200
198 202
199 to 201
300
298 302
299 to 301
* Dyno-lube
J22 of dyno
23
Tdyn_aft
32
30 34
31 to 33
Aft Temp.
100
98 102
99 to 101
200
198 202
199 to 201
300
298 302
299 to 301
* Dyno-lube
J20,J23,
none
Tdyn_FA
32
30 34
31 to 33
Temp.FWD/AFT
See note(below)
100
98 102
99 to 101
Average
200
198 202
199 to 201
300
298 302
299 to 301
* The Limit Control Panel (paragraph #37) may be calibrated in conjunction with the channels
marked by "*".
NOTE
To align a Dyno-lube delta temperature (above), disconnect J24, install thermocouple at
J20, and Connect Calibrator at J23. Record the observed temperature. Then install
calibrator at J24 and open J23 (leave thermocouple at J20) and the output should be same
as recorded above.
26