TB 1-4920-443-35
(6) Set the thermocouple calibrator for an output of 1000 F.
(7) Verify that the TI temperature indicator indication is between 998F and 1002 F. Adjust the indicator
high potentiometer as necessary (R15 for digital, and R21 for analog, high adjustments).
NOTE
Interaction may occur between the TI temperature indicator low and high adjustments.
Repeat steps 4 through 7 until no further adjustment is necessary.
(8) Adjust the calibrator output to obtain the temperature values listed below. Record each temperature
indication and verify that these values are within the tolerance limits listed in the table below.
Table 33-1 TI Temperature Indications
Thermocouple Temperature Indicator Actual TI Reading
Calibrator (F)
Tolerance Limits
DIGITAL
400
398 to 402
800
798 to 802
1200
1198 to 1202
1600
1598 to 1602
1650
1648 to 1652
2000
1998 to 2002
29