TM 5-6675-319-14
d. With a screwdriver, turn the thermostat adjusting screw. Turn to right
to raise, to left to lower temperature. Glow lamp will light when the
developer heater is energized. Monitor DEVELOPER thermometer. Tempera-
ture indicated should be 80°F (26.6°C).
e.
Check temperature with self-indicating bimetallic thermometer.
f . R e i n s t a l l d r i p c a p.
NOTE
If developer temperature other than 80°F (26.6°C) is required, recycling
tank thermostat must be set to 5°F (3°C) less than required developer
temperature before being adjusted.
g.
If DEVELOPER thermometer reading still does not match the bimetallic
thermometer reading of 80°F (26.6°C), the DEVELOPER thermometer must be
adjusted (paragraph 2-16.4).
h .
R e i n s t a l l r e a r a c c e s s c o v e r.
2-99