d. Remove
e .
Remove
f. Remove
g.
Remove
h . L o o s en
TM 5-6675-319-14
thumbscrew and remove flowmeter shield.
retaining nuts securing flowmeter to processor housing.
screws from upper manifold.
retaining screw and upper manifold from assembly.
hex head screws and remove flowmeter tubes from housing. Remove
stops and ball from tube.
i .
Clean flowmeter tube with warm water and tube cleaning brush.
j .
Install new ball and stops in flowmeter tubes.
k. Remove control valves from flowmeter.
1.
Remove O-rings and washers from control valves, and replace.
m.
R e i n s t a l l c o n t r o l v a l v e s i n f l o w m e t e r.
n .
Reinstall upper manifold and retaining screws.
o .
Remove flowmeter assembly on processor frame. Then reinstall flowmeter
tubes.
p.
Reinstall shield and tighten thumbscrew.
q.
Reconnect input
r .
Check flowmeter
s.
R e i n s t a l l r i g h t
t .
Turn on power.
and output lines.
flow rates (Table 2-1, PMCS item 19).
side panel.
2-103