TM 5-6675-323-14
(8) Place linear measuring probe and stand so that pricking pin is at
beginning mark of line.
(9) Release foot switch.
(10) Grasp probe body and roll linear measuring probe along length of line.
Stop when needle is precisely at end of line.
(11) Press foot switch.
(12) Divide digital readout on LED display by 10 to determine alined pulse
c o u n t r e a d i n g , RL .
(13) Record this number.
(14) Repeat steps (6) through (13) four times.
(15) Average the five pulse count results obtained from the LED display after
e a c h test measurement to obtain R L v a l u e .
NOTE
I f a v e r a g e RL v a l u e i s n o t w i t h i n 0 . 2 % o f f a c t o r y d e t e r m i n e d
R L value, the linear measuring probe may be defective. Repeat
alinement procedure. Notify direct/general support maintenance
for
service or replacement.
( 1 6 ) C o m p a r e a v e r a q e RL, value to R L v a l u e f o u n d o n C a l i b r a t i o n R e c o r d s S h e e t
provided-with equipment.
NOTE
I f u s i n g preprogrammed c a r d , P r o g r a m I I I , m a n u a l c a l c u l a t i o n o f CL i s
not necessary.
( 1 7 ) Use scale of map, plan, or blueprint being measured to determine C L
( c a l i b r a t i o n c o n s t a n t ) . C L is needed to perform various measurements.
( 1 8 ) U s e t h e f o l l o w i n g e q u a t i o n t o d e t e r m i n e CL .
= S/R L
CL
S = Scale, in desired units for measurements.
can be found on Calibration Records Sheet or from average of test
RL
measurements.
9-54