TM 5-6675-319-14
2 - 1 0 . 4 S e r v i c e S t r a i n e r s .
MOS:
83E, Photo and Layout Specialist
SUPPLIES:
Strainer Gaskets
P a i l
a.
Turn off all processor switches and RECYCLING PUMP switch.
b.
c.
d.
e.
f .
g.
h.
i .
j .
Remove left side panel.
Drain developer and fixer tanks by opening DEV and FIX drain valves.
Disconnect developer and fixer replenishment lines at quick-disconnects.
Unscrew strainer caps. Remove gaskets and screens. Discard gaskets.
Wash screens in warm water.
Install screens, new gaskets, and caps on strainer bodies.
Replace side panel.
Reconnect developer and fixer replenishment lines.
Refill developer and fixer tanks, and check for leaks.
2-62