Table 2-3. ORGANIZATIONAL PREVENTIVE MAINTENANCE CHECKS AND SERVICES
TM 5-6675-319-14
NOTE
If the equipment must be kept in continuous operation, check and service
only those items that can safely be checked and serviced without distur-
bing operation. Make the complete checks and services when the equipment
can be shut down.
B- Before
W- Weekly
AN - Annually
(Number) - Hundreds of Hours
D - During
M - Monthly
S
- Semiannually
A - After
ITEM
NO.
1
IN-
TER-
VAL
M
Q - Quarterly
BI
- Biennially
ITEM TO BE INSPECTED
PROCEDURE
FILM/PAPER PROCESSOR SYSTEM
Adjust Developer Temperature.
CAUTION
Do not use mercury-filled thermometer to check
temperature. Accidental breakage could result
in serious mercury contamination of unit.
1. Adjust water temperature at the recycling tank
thermostat to maintain a temperature of 75°F
(23.8°C).
2. Turn on DRIVE, PUMP, and DRYER switches.
3.
Set DRYER THERMOSTAT to 110°F (43.3°C) and FILM
TRANSPORT SPEED indicator to 30 in. per minute.
4. Run unit for at least 30 minutes to allow developer
solution to reach the recommended operating tem-
perature of 80°F (26.6°C).