Table 4-3. ORGANIZATIONAL TROUBLESHOOTING
MALFUNCTION
TEST OR INSPECTION
CORRECTIVE ACTION
1.
DEISITOMETER WILL NOT ENERGIZE.
Step. 1.
Check for power at outlet.
(a)
If power present, proceed to step 2.
(b)
Restore power to outlet.
Step. 2.
Check for defective power cord.
(a)
If power cord is not defective, proceed to step 3.
(b)
Replace power cord (paragraph 4-16.11).
Step. 3.
Check for defective power switch.
(a)
If power switch is not defective, proceed to step 4.
(b)
If power switch defective, replace switch. (paragraph 4-16.9)
Step.
4.
Check for defective fuse holder.
(a)
If fuse holder is not defective, proceed to step 5.
(b)
If fuse holder is defective, replace fuse holder. (paragraph 4-16.6)
Step. 5.
Check for defective power select switch.
(a)
If power select switch is not defective, proceed to step 6.
(b)
If power select switch is defective, replace power select switch. (paragraph 4-16.5)
Step. 6.
Check for defective power transformer.
(a)
If power transformer is not defective, proceed to step 7.
(b)
If power transformer is defective, replace transformer. (paragraph 4-16.8)
4-17