TM 5-3610-257-14
4-6.4.
Preparation for Movement.
a.
Set densitometer POWER switch to OFF position.
b.
Unplug reflection probe and place probe in secure storage.
4-7.
OPERATION UNDER UNUSUAL CONDITIONS. This equipment is designed for operation only in a controlled
environment.
Section III. OPERATOR MAINTENANCE
4-8.
LUBRICATION INSTRUCTIONS. This equipment does not require lubrication.
4-9.
TROUBLESHOOTING PROCEDURES.
a.
The table lists the common malfunctions which you may find during operation or maintenance of the
densitometer, or its components. You should perform the tests/inspections and corrective actions in the order listed.
b.
This manual cannot list all malfunctions that may occur, nor all test or inspections and corrective actions.
If a malfunction is not listed or is not corrected by listed corrective actions, notify your supervisor.
Table 4-2. TROUBLESHOOTING
MALFUNCTION
TEST OR INSPECTION
CORRECTIVE ACTION
1.
DENSITOMETER WILL NOT TURN ON.
Step 1.
Check that power cord is plugged in and circuit breaker is on
(a)
Plug in power cord
(b)
Turn circuit breaker on
(c)
If no power at duplex receptacle, refer to higher level of maintenance.
Step 2.
Check for defective fuse.
4-14