TM 5-3610-257-14
d.
Adjust ZERO control knob of channel selected until a reading of 0.000 is obtained.
e.
Repeat step d. for each channel position.
f.
Place cyan-colored cali-button in probe tip.
g.
Place CHANNEL control switch in the cyan-colored position.
h.
Press READ pushbutton.
i.
If necessary, adjust SLOPE control so that display shows reading recorded on cali-button case top.
j.
Repeat step i.
for each of the remaining colors.
4-6.3.
Operating Procedures.
a.
Place probe on unknown area of sample or copy.
b.
Center desired area in probe target.
c.
Press upper portion of probe downward until it rests firmly on probe base.
d.
Depress and hold READ pushbutton until a reading is obtained on the digital readout. If desired, move
probe or sample around, while holding the READ pushbutton to obtain maximum and minimum values.
NOTE
The digital readout will automatically hold and display the last measured value when the READ
pushbutton is released.
e.
Release READ pushbutton when desired reading is obtained.
f.
Return hinged portion of probe to its raised position.
4-13