TM 5-3610-257-14
4-6.
OPERATION UNDER USUAL CONDITIONS.
4-6.1.
Assembly and Preparation for Use.
CAUTION
Be sure correct voltage is selected prior to applying power or serious damage to equipment may
occur.
a.
Be sure densitometer voltage selection switch on rear panel is set to correct position for the power
available.
b.
Plug in reflection probe cable.
c.
Plug in densitometer.
NOTE
For the highest possible accuracy, the densitometer should be turned on and allowed to reach
normal operating temperature (one half hour or more) before use.
d.
Move POWER switch to ON position.
e.
Be sure digital display and reflection probe lamp light.
4-6.2.
Initial Adjustments, Daily Checks, and Self Test.
a.
Turn on densitometer and allow to warm up for one half hour.
b.
Place white cali-button in probe tip.
c.
Press READ pushbutton.
4-12