TM 5-6675-319-14
Table 2-4.
ORGANIZATIONAL TROUBLESHOOTING - Cont
MALFUNCTION
TEST OR INSPECTION
CORRECTIVE ACTION
5.
AFTER WARM-UP, DEVELOPER THERMOMETER DOES NOT READ 80°F (26.6°C) - Cont
CAUTION
Do not use mercury-filled thermometer to check temperature. Breakage
may result in serious mercury contamination of unit.
( b ) I n s e r t t h e r m o m e t e r i n t o d e v e l o p e r s o l u t i o n t a n k.
(c) Compare DEVELOPER thermometer reading with reading on
thermometer in solution tank.
(d) Adjust DEVELOPER thermometer if reading varies (paragraph
2-16.4).
(e) If DEVELOPER thermometer and thermometer in solution tank are
the same temperature, proceed to step 2.
Step 2.
Check for improperly adjusted developer heater thermostat while
processor is activated.
2-81