TM 5-6675-319-14
Table 2-4. ORGANIZATIONAL TROUBLESHOOTING - Cont
MALFUNCTION
TEST OR INSPECTION
CORRECTIVE ACTION
5.
AFTER WARM-UP, DEVELOPER THERMOMETER DOES NOT READ 80°F (26.6°C) - Cont
NOTE
Adjustment shall be performed with DRYER THERMOSTAT set at 11O°F (43.3°C)
and FILM TRANSPORT SPEED indicator set at 30 in. per minute.
(a) Remove retaining screws and rear cover.
(b) Remove drip cap from thermostat housing.
(c) Check glow lamp for activation which indicates heater element
i s a c t i v a t e d.
Adjust DEVELOPER thermostat (paragraph 2-16.6).
2-82