TM 5-6675-317-14
Table 5-2. ORGANIZATIONAL TROUBLESHOOTING - Cont
MALFUNCTION
TEST OR INSPECTION
CORRECTIVE ACTION
2. NO WATER AGITATION.
Step 1.
Using multimeter, check for continuity of power cord.
( a ) I f c o n t i n u i t y e x i s t s , p r o c e e d t o s t e p 2 .
(b) If continuity does not exist, replace power cord (paragraph
5-16.1).
Step 2.
Check continuity of power switch.
(a) If continuity does not exist, replace power switch (paragraph
5-16.2).
(b) If continuity does exist, replace circuit board
(paragraph 5-16.3).
5-16. MAINTENANCE PROCEDURES.
a. This section contains instructions covering organizational maintenance
functions for the ultrasonic cleaner.
Personnel required are listed only if the
task requires more than one.
b.
After completing each maintenance procedure, perform operational check to be
sure that equipment is properly functioning.
INDEX
PROCEDURE
PARAGRAPH
Replace Power Cord . . . . . . . . . . . . . . . . . . . . . . . . . .
5-16.1
Replace Power Switch . . . . . . . . . . . . . . . . . . . . . . . . .
5-16.2
Replace Circuit Board . . . . . . . . . . . . . . . . . . . . . . . .
5-16.3