TM 5-6675-323-14
Table 313. TROUBLESHOOTING - Cont
MALFUNCTION
TEST OR INSPECTION
CORRECTIVE ACTION
EVERYTHING ELSE IS INOPERATIVE - Cont
3.
COOLING FAN RUNS .
NOTE
T h i s a c t i o n r e s e t s p o w e r s u p p l y i f f a u l t was caused by momentary l o w l i n e
voltage.
PLATEN DOES NOT HOLD PAPER WHEN "CHART H O L D ' IS ACTIVATED.
4.
C h e c k that paper is not wet and platen i s c l e a n .
(1) Clean platen if necessary.
(2) Replace paper.
NO RESPONSE TO PROGRAMMED COMMANDS.
5.
S t e p 1.
Check to see if front panel ERROR lamp is lit.
(a) If lamp is not lit, proceed to step 2.
(b) If lamp is on, correct improper program command.
Check that HP-IB Interface is connected and tight.
S t e p 2.
(a) If HP-IB Interface is properly connected, proceed to step 3.
(b) Connect and tighten HP-IB Interface.
S t e p 3.
Check that HP-IB Interface select code switch is set to 7.
(a) If code switch is properly set, proceed to step 4.
(b)
Turn
off
controller's
power.
(c) Set switch to 7.
(d) Re-energize controller.
S t e p 4.
Refer HP-IB to vendor for servicing.
3-112