TM 5-3610-285-14
WARNING
Death or serious injury may occur from electrical shock unless power cord is
unplugged before servicing equipment.
NOTE
Aperture assembly should remain in place on the clear aperture support tube.
e. Remove table platen.
NOTE
The optical system is alined by tilting the lamp mounting block and/or rotating
the mirror support bracket.
f.
If mirror support must be turned, temporarily insert a screw into the
pressed-in nut to prevent unintentional turning; then turn mirror
support to desired position.
g.
If bracket support must be turned, turn bracket to desired position.
h.
Secure light table platen with screws.
i.
Recheck lamp alinement.