TM 5-3610-285-14
Table 7-2. TROUBLESHOOTING - Cont
MALFUNCTION
TEST OR INSPECTION
CORRECTIVE ACTION
2.
TRANSMISSION PROBE LIGHT TABLE WILL NOT LIGHT.
Step 1.
Check that power cord is plugged in.
(a) If power cord is plugged in, proceed to step 2.
(b) Connect power cord.
Step 2.
Check position of power panel circuit breaker.
(a) If circuit breaker is on, refer to step 3.
(b) Reset circuit breaker.
WARNING
Death or serious injury may occur from electrical shock unless power cord is unplugged before
troubleshooting equipment.
Step 3.
Check for blown fuse.
(a) If fuse is not blown, notify supervisor.
(b) Change fuse.
7-10.
MAINTENANCE PROCEDURES . There are no operator maintenance procedures assigned for this equipment.
7-11.
LUBRICATION INSTRUCTIONS . This equipment does not require lubrication.
7-21