TM5-6675-326-14
4-14.3 Replace Power Cord - Continued
REMOVAL:
WARNING
ELECTRICAL SHOCK
Unplug power cord before servicing the
zoom transfer scope.
Failure to do so
may result in death or serious injury.
a .
Turn power OFF, and unplug power cord.
b.
Unplug below-stage illuminator cord.
c.
Remove fuse caps and fuses.
d.
Remove six screws from cover on underside of crosspiece
assembly.
e .
Release strain relief bushing from power cord.
f .
Push power cord into crosspiece assembly to allow for give.
g.
Note color of wires connected to fuse holders and to ground.
Disconnect wires from fuse holders and ground.
h .
Remove power cord through hole in crosspiece assembly.
INSTALLATION:
a .
Install new power cord through hole in crosspiece assembly.
b.
Connect wires to ground and fuse holders.
c . R e i n s t a l l s t r a i n r e l i e f b u s h i n g o n p o w e r c o r d .
d.
Reinstall cover on underside of crosspiece assembly, and
secure with six screws.
e.
Reinstall fuses and fuse caps.
f .
Plug in below-stage illuminator cord.
g.
Plug in power cord , and turn power ON.
4-38