TM 5-6675-316-14
Table 7-3. ORGANIZATIONAL TROUBLESHOOTING
MALFUNCTION
TEST OR INSPECTION
CORRECTIVE ACTION
1.
LAMPS DO NOT LIGHT.
Step 1.
Check continuity of power switch.
(a) If continuity exists, proceed to step 2.
(b) If no continuity exists, replace power switch (paragraph
7-16.1).
Step 2.
Check continuity of power cord.
(a) If no continuity exists, replace power cord (paragraph
7-16.2).
(b) If continuity exists, replace tube starter (paragraph 7-16.5).
(c) If lamps still do not light, replace ballast (paragraph
7-16.4).
2.
POWER RECEPTACLES DO NOT WORK.
Step 1.
Check continuity of power cord.
(a) If continuity exists, proceed to step 2.
(b) If no continuity exists, replace power cord (paragraph
7-16.2).
Step 2.
Check continuity of receptacle.
Repair receptacle (paragraph 7-16.3).
3.
TABLE DOES NOT LOCK.
Step 1.
Check for loose tilt lock.
(a) If tight, proceed to step 2.
(b) Tighten tilt lock.
7-17