TM 5-3610-285-14
Table 7-3. ORGANIZATIONAL TROUBLESHOOTING - Cont
MALFUNCTION
TEST OR INSPECTION
CORRECTIVE ACTION
1. DENSITOMETER WILL NOT ENERGIZE - Cont
(b) If power capacitor defective, replace capacitor (paragraph 7- 16.13).
2. REFLECTION PROBE LAMP WILL NOT LIGHT.
Step 1. Check for defective lamp.
(a) If lamp not defective, proceed to step 2.
(b) If lamp defective, replace lamp (paragraph 7-16.2).
Step 2. Check for broken wires or connection in probe.
(a) If wires and connections are intact, proceed to step 3.
(b) Replace/repair any loose or broken wires.
Step 3. Check for defective probe connector.
(a) If probe connector not defective, notify supervisor.
(b) If probe connector defective, replace connector (paragraph 7- 16.14).
3. NO RESPONSE TO PROBE READ PUSHBUTTON.
Step 1. Check for defective read pushbutton.
(a) If pushbutton not defective, proceed to step 2.
(b) If pushbutton defective, replace pushbutton (paragraph 7-16.4).
Step 2. Check for defective probe connector.
(a) If connector not defective, notify supervisor.
(b) If connector defective, replace connector (paragraph 7-16.14).
7-30