TM 5-3610-285-14
Table 7-3. ORGANIZATIONAL TROUBLESHOOTING
MALFUNCTION
TEST OR INSPECTION
CORRECTIVE ACTION
1. DENSITOMETER WILL NOT ENERGIZE.
Step 1. Check for power at outlet.
(a) If power present, proceed to step 2.
(b) Restore power to outlet.
Step 2. Check for defective power cord.
(a) If power cord is not defective, proceed to step 3.
(b) Replace power cord (paragraph 7-16.11).
Step 3. Check for defective power switch.
(a) If power switch is not defective, proceed to step 4.
(b) If power switch defective, replace switch (paragraph 7-16.5).
Step 4. Check for defective fuse holder.
(a) If fuse holder is not defective, proceed to step 5.
(b) If fuse holder defective, replace fuse holder (paragraph 7- 16.6).
Step 5. Check for defective power select switch.
(a) If power select switch is not defective, proceed to step 6.
(b) If power select switch defective, replace power select switch (paragraph 7-16.9).
Step 6. Check for defective power transformer.
(a) If power transformer is not defective, proceed to step 7.
(b) If power transformer defective, replace transformer (paragraph 7-16.8).
Step 7. Check for defective power capacitor.
(a) If power capacitor is not defective, notify supervisor.
7-29