TM5-6675-326-14
7-13.1 Replace Power Cord - Continued
REMOVAL:
WARNING
ELECTRICAL SHOCK
Unplug power cord before servicing the ultrasonic
cleaner. Failure to do so may result in death or
s e r i o u s i n j u r y .
a .
Turn power OFF.
Unplug power cord.
b .
Remove screws and nuts holding stainless steel tank and casing
t o c h a s s i s.
c .
L i f t s t a i n l e s s s t e e l t a n k a n d c a s i n g f r e e.
Set aside.
NOTE
Do not disconnect wires to transducer.
d.
Remove screws and nuts holding circuit board to chassis.
e.
Disconnect power cord wire from power switch.
f .
Disconnect power cord wire from circuit board.
g.
Remove defective power cord from chassis.
h.
Remove strain relief bushing from defective peer cord.
INSTALLATION:
a .
put strain relief bushing on new power cord.
Put terminal
ends of cord into chassis.
b.
Fit strain relief bushing into power cord hole.
c . Reconnect power cord wire to circuit board.
d.
Reconnect power cord wire to power switch.
e .
Secure circuit board to chassis with screws and nuts.
f .
Reinstall stainless steel tank and casing.
Secure with nuts
and screws.
g .
F i l l s t a i n l e s s s t e e l t a n k 1 / 3 f u l l w i t h w a t e r.
h .
Plug in power cord. Turn power ON.
Be sure water surface agitates.
7-11