TM 5-6675-323-14
a.
Turn off power switch and unplug power cord.
b. Unplug above-stage and table illuminator cords.
c.
R e m o v e screws from cover on underside of crosspiece assembly. Move
cover to allow access to potentiometer.
d. Loosen setscrews on illuminator control knob. Remove knob, nut, and
washer.
e.
Remove potentiometer through underside of crosspiece assembly.
f. Loosen screw on potentiometer and remove mount.
Tag and desolder wires.
g.
h. Solder wires to new potentiometer.
i.
Reinstall mount and tighten screw.
Position potentiometer in crosspiece assembly.
j.
Reinstall knob and tighten setscrews.
k. Reinstall washer and hex nut.
l.
Reinstall cover on underside of crosspiece assembly and secure with
screws.
m.
Plug in above-stage and table illuminator cords.
P l u g in power cord and turn on power.
n.
8-45