TM 5-6675-319-14
Table 2-4. ORGANIZATIONAL TROUBLESHOOTING - Cont
MALFUNCTION
TEST OR INSPECTION
CORRECTIVE ACTION
10.
WATER IN RECYCLING TANK DOES NOT MAINTAIN TEMPERATURE OF THERMOSTAT SETTING
WHEN ACTIVATED - Cont
Step 3.
Check thermostat for proper operation as follows:
(a) Remove retaining screws and heating element cover.
WARNING
Electrical shock hazard. You must stand on rubber matting as a protective
measure while performing this procedure. Death or serious injury may occur.
(b) Adjust thermostat setting above the temperature indicated by the black
needle and check for voltage at the heating element.
(1) Replace thermostat if 110 V ac is not present (paragraph
2 - 1 6 . 2 0 ) .
(2) If 110 V ac is present, replace heating element (paragraph
2 - 1 6 . 1 9 .
2-16. ORGANIZATIONAL MAINTENANCE PROCEDURES.
a . T h i s s e c t i o n c o n t a i n s i n s t r u c t i o n s c o v e r i n g o r g a n i z a t i o n a l m a i n t e n a n ce
functions for the film/paper processor system. Personnel required are listed only
if the task requires more than one.
b. After completing each maintenance procedure, perform operational check to be
sure that equipment is properly functioning.
2-88