TM 5-3610-252-14
Table 3-3. ORGANIZATIONAL TROUBLESHOOTING - Cont
MALFUNCTION
TEST OR INSPECTION
CORRECTIVE ACTION
HUMIDITY INDICATOR NEEDLE REMAINS DEFLECTED AND CANNOT
- Cont
j.
Be sure that scribed line on null control knob
BE ZEROED WITH KNOB
indicates relative
humidity of 47% ± 1%.
NOTE
If reading is not correct, remove cable and
reading is correct, cable is bad. If still
Replace either reader or cable according to
connect reader test plug. If
incorrect, reader is bad.
results of test.
3-16. MAINTENANCE PROCEDURES.
There are no organizational maintenance procedures
assigned for this equipment.
3-17. PREPARATION FOR STORAGE OR SHIPMENT.
Contact your battalion for packing
and shipping instructions.
Section V DIRECT/GENERAL SUPPORT MAINTENANCE
There are no assigned direct/general support maintenance tasks for this equipment.
3-21/(3-22 blank)