TM 5-3610-252-14
Table 3-3. ORGANIZATIONAL TROUBLESHOOTING - Cont
MALFUNCTION
TEST OR INSPECTION
CORRECTIVE ACTION
HUMIDITY INDICATOR NEEDLE REMAINS DEFLECTED AND CANNOT BE ZEROED WITH KNOB
- Cont
f. Scribed line on null control knob should indicate temperature of between
98° and 102°F.
NOTE
If reading is not correct, remove cable and
Repeat above steps.
If reading is correct,
g.
Slide TEMP/RH switch down to RH.
h. Rotate DIAL D until arrow on reader
COMPENSATION scale.
connect reader test plug.
replace cable.
panel points to 100°F on
i.
Rotate null control knob until HUMIDITY INDICATOR needle points at center
point.
3-20