to be calibrated.
(2) Perform the following Calibration Procedures.
(3) Select the channel to be calibrated and double click. The operator will be prompted to enter the 1st
(4) Set the signal source to 1st Point (Low Value) and then click the Set 1st Point button. Refer to Figure
4-15 through Figure 4-19 for Test Set-ups.
(5) Set the signal source to 2nd Point (High Value) and then click the Set 2nd Point button. The Two
Point Calibration tab will display "Two point calibration complete on T_nnnnn.", where T_nnnnn is the
name of the transducer.
(6) Click OK and the following Confirmation screen will be displayed
Figure 7. Set-Point Overwrite Confirmation Screen
(7) Click Yes to
overwrite the existing transducer calibration and return to
the TCIP Executive window.
Click No to not overwrite the existing transducer calibration and return to the Calibration screen.
d. MANUAL CALIBRATION
A manual calibration consists of entering known gain and offset values for a channel.
(1)
Perform the procedures of Section C.
(2)
Select the channel to be calibrated and double click.
(3)
Click the Manual Calibration tab.
(4)
Enter gain value (slope) and offset value (intercept) as shown in Figure 8.
93